About this example
When the spread of the differences grows with the size of the measurement, as is common for laboratory analytes over a wide range, percentage differences, 100 × (A − B) divided by the mean, keep the limits meaningful across the range. These 20 pairs of glucose values (mmol/L) are illustrative, not from a study.
The bias is −0.1% and the 95% limits of agreement are −9.8% to 9.6%, inside the ±15% acceptable limits that were entered as a pre-specified criterion. The tool only draws those limits; deciding them, and judging the result, belongs to the study protocol and the relevant standard.
The data
| ID | Point of care | Laboratory |
|---|---|---|
| G01 | 4.2 | 4.4 |
| G02 | 5.1 | 5 |
| G03 | 5.6 | 5.9 |
| G04 | 6.3 | 6.1 |
| G05 | 7.4 | 7 |
| G06 | 8.1 | 8.4 |
| G07 | 9 | 8.6 |
| G08 | 10.2 | 10.6 |
| G09 | 11.6 | 11 |
| G10 | 12.9 | 13.5 |
| G11 | 4.8 | 5 |
| G12 | 6 | 5.8 |
| G13 | 6.9 | 7.2 |
| G14 | 7.7 | 7.3 |
| G15 | 8.8 | 9.3 |
…and 5 more rows. Open the example to see all of them.
Illustrative paired values. Differences are percentages of the mean of the two methods; limits are bias ± 1.96 SD. Open the example, replace the data with your own, and export the figure as SVG, PDF, PNG or TIFF.