Respiratory physiology (Bland & Altman 1986)

Bland–Altman example: two peak flow meters

The classic Bland–Altman example: Wright vs mini Wright peak flow in 17 people. Bias −2.1 l/min, 95% limits of agreement −78 to 74 l/min.

Open this example in the editorStart from scratch

−100−50050100200300400500600700Mean of Wright and Mini Wright, l/minWright − Mini Wright, l/min+1.96 SD 73.9Mean −2.1−1.96 SD −78.1
Wright minus mini Wright peak expiratory flow (l/min) against their mean for 17 people; solid line = bias, dashed lines = 95% limits of agreement.

About this example

These are the data from Bland and Altman’s 1986 Lancet paper that introduced the plot: peak expiratory flow rate measured on 17 people with a large Wright peak flow meter and a mini Wright meter. The example uses the first reading on each meter. The two readings correlate strongly (r ≈ 0.94), yet that says nothing about whether one meter can replace the other.

The mean difference is −2.1 l/min (95% CI −22.0 to 17.8), so there is no clear systematic bias. The SD of the differences is 38.8 l/min, giving 95% limits of agreement of −78.1 to 73.9 l/min: for most people the mini meter may read up to about 74 l/min below or 78 l/min above the large meter. Whether that is acceptable is a clinical question, decided before looking at the data.

Sources: Bland & Altman (1986), Statistical methods for assessing agreement between two methods of clinical measurement; BlandAltmanLeh: bland.altman.PEFR data documentation

The data

SubjectWrightMini Wright
1494512
2395430
3516520
4434428
5476500
6557600
7413364
8442380
9650658
10433445
11417432
12656626
13267260
14478477
15178259

…and 2 more rows. Open the example to see all of them.

One row per subject, first reading on each meter (Bland & Altman 1986). Bias, limits of agreement and their 95% CIs match base R and BlandAltmanLeh. Open the example, replace the data with your own, and export the figure as SVG, PDF, PNG or TIFF.