About this example
These are the data from Bland and Altman’s 1986 Lancet paper that introduced the plot: peak expiratory flow rate measured on 17 people with a large Wright peak flow meter and a mini Wright meter. The example uses the first reading on each meter. The two readings correlate strongly (r ≈ 0.94), yet that says nothing about whether one meter can replace the other.
The mean difference is −2.1 l/min (95% CI −22.0 to 17.8), so there is no clear systematic bias. The SD of the differences is 38.8 l/min, giving 95% limits of agreement of −78.1 to 73.9 l/min: for most people the mini meter may read up to about 74 l/min below or 78 l/min above the large meter. Whether that is acceptable is a clinical question, decided before looking at the data.
Sources: Bland & Altman (1986), Statistical methods for assessing agreement between two methods of clinical measurement; BlandAltmanLeh: bland.altman.PEFR data documentation
The data
| Subject | Wright | Mini Wright |
|---|---|---|
| 1 | 494 | 512 |
| 2 | 395 | 430 |
| 3 | 516 | 520 |
| 4 | 434 | 428 |
| 5 | 476 | 500 |
| 6 | 557 | 600 |
| 7 | 413 | 364 |
| 8 | 442 | 380 |
| 9 | 650 | 658 |
| 10 | 433 | 445 |
| 11 | 417 | 432 |
| 12 | 656 | 626 |
| 13 | 267 | 260 |
| 14 | 478 | 477 |
| 15 | 178 | 259 |
…and 2 more rows. Open the example to see all of them.
One row per subject, first reading on each meter (Bland & Altman 1986). Bias, limits of agreement and their 95% CIs match base R and BlandAltmanLeh. Open the example, replace the data with your own, and export the figure as SVG, PDF, PNG or TIFF.